翻訳と辞書
Words near each other
・ Integrated design
・ Integrated Design Associates
・ Integrated development environment
・ Integrated device manufacturer
・ Integrated Device Technology
・ Integrated digital platform
・ Integrated digital television
・ Integrated Diver Display Mask
・ Integrated DNA Technologies
・ Integrated Dynamics
・ Integrated education
・ Integrated electric propulsion
・ Integrated Electrical Services
・ Integrated Electronic Control Centre
・ Integrated Electronic Litigation System
Integrated electronic piezoelectric accelerometer
・ Integrated Encryption Scheme
・ Integrated engineering
・ Integrated enterprise modeling
・ Integrated Facility for Linux
・ Integrated Farm Management Program
・ Integrated Farm School
・ Integrated farming
・ Integrated financial advisor (India)
・ Integrated floating cage aquageoponics system
・ Integrated flowthrough
・ Integrated fluidic circuit
・ Integrated fluorometer
・ Integrated Food Security Phase Classification
・ Integrated Forecast System


Dictionary Lists
翻訳と辞書 辞書検索 [ 開発暫定版 ]
スポンサード リンク

Integrated electronic piezoelectric accelerometer : ウィキペディア英語版
Integrated electronic piezoelectric accelerometer

Integrated electronic piezoelectric (IEPE) accelerometers are a class of piezoelectric accelerometers that incorporate an electronic amplifier and use a single two-pole coaxial connector for both power input and signal output.〔Jon S. Wilson, ''Sensor technology handbook'', Newnes, ISBN 0-7506-7729-5, 2004, p.289〕 The devices are also known by proprietary names such as integrated circuit piezoelectric sensor (ICP).〔Neil J. Mansfield, ''Human response to vibration'', 1st edn., CRC Press, 2004, ISBN 0-415-28239-X, p.106〕 Compared with earlier sensor interface systems requiring a charge-sensitive preamplifier external to the sensor, the use of a single coaxial connector with integrated electronics gives IEPE devices their main advantages: small size, light weight, robustness and low cabling costs. Microphones and other vibration sensors are also made using an IEPE-compatible interface.
== Electrical characteristics ==
The IEPE interface is defined in the IEEE 1451.4 standard〔http://standards.ieee.org/regauth/1451/index.html〕 which also includes the definition for the Transducer Electronic Data Sheet (TEDS), data from which is carried on the same interface. Some IEPE sensors are made without TEDS, using only the powering interface defined by IEEE 1451.4. Different manufacturers' implementations vary, so the manufacturer's data sheet must be consulted for each device.〔(''Practical considerations in using IEPE accelerometers with modern data acquisition systems'' ), Meggitt/Endevco 2009, accessed 2009-05-18〕
Power is supplied to the inner conductor of the coaxial cable from an external constant-current supply of 0.5 to 8 mA. The power supply can be built into the receiving instrument such as an oscilloscope or data acquisition device, or may be a separate unit.
The output signal from the device is also on the inner conductor, and consists of an AC voltage 〔("What is the output of an IEPE accelerometer?" ), Honeywell Test & Measurement, accessed 2009-04-14〕 centred on a bias voltage of about 8 to 12 V.〔("IEPE Standard" ), Metra Mess- und Frequenztechnik in Radebeul e.K., accessed 2009-05-18〕 The output voltage is bounded at the lower end by the saturation voltage of the integrated IEPE amplifier (typically 0.5 to 2 volts), and at the upper end by the maximum compliance voltage of the current source (which may be anything between about 6 and 30 volts).
The shield of the cable serves as the ground return for the supply current and also protects the inner conductor from electromagnetic interference.
This method of combining the power and signal on a single pair of conductors is a form of simplex powering, sometimes referred to as phantom although it is not compatible with the phantom powering system used in audio equipment. The receiving instrument uses a simple RC filter to separate the output signal of the IEPE device from the supply current. This means that the DC component of the signal is lost, and so when a static acceleration needs to be measured, a "zero Hz" accelerometer must be used instead.